Vapour Phase or Liquid Phase? Choosing the Right Way to Store Samples

Cryonos06-1920-1280x717.webp
Kriyojenik depolama laboratuvarı
Cryonos

Vapour Phase or Liquid Phase? Choosing the Right Way to Store Samples

Cryogenic storage in liquid nitrogen (LN2) is the standard way to preserve the viability of cells, tissue, cord blood and other biological samples for many years. But one of the first and most important decisions when choosing a freezer is whether samples will be stored immersed in liquid nitrogen (liquid phase) or in the cold nitrogen vapour above the liquid (vapour phase). Both approaches have strengths and points that need attention.

Why below −135 °C?

The point at which biological activity in cells is considered to stop in practice is the glass transition temperature of water, about −135 °C. Below this temperature ice crystal growth and chemical reactions become negligible. The goal in long-term storage is therefore to keep samples clearly below this limit at all times. Many laboratories use −150 °C as a safety margin.

Liquid-phase storage

Samples are immersed directly in liquid nitrogen at −196 °C.

  • Advantages: The lowest and most stable temperature; a large buffer against fluctuations when the lid is opened or during filling; a long safe period during a power cut.
  • Points to watch: Liquid nitrogen is not sterile, and cross-contamination between samples via liquid nitrogen has been reported in the literature. Liquid nitrogen can seep into leaking cryovials and expand rapidly on thawing, causing the vial to burst. Suitable cryovials, a second layer of protection (overwrap) and freezing bags with verified seal integrity are therefore especially important for samples stored in the liquid phase.
Cryonos cryogenic solutions (illustrative image)
Cryonos cryogenic solutions (illustrative image)

Vapour-phase storage

Samples are kept in the cold vapour zone above the liquid nitrogen. In conventional designs the liquid nitrogen sits at the bottom of the chamber; in jacketed designs it circulates within the chamber walls and the sample compartment stays completely dry.

  • Advantages: Because samples do not touch liquid nitrogen, the risk of cross-contamination is greatly reduced; liquid ingress into vials, and the related bursting risk, is eliminated; access to samples is safer and more ergonomic.
  • Points to watch: A vertical temperature gradient can form inside the chamber, with the upper zone warmer than the lower zone. Temperature rises faster when the lid is left open or with frequent access. The liquid nitrogen level must be kept within the range the design intends, and temperature should be monitored, especially in the upper zone. A well-designed vapour-phase freezer can keep even the top level below −150 °C, depending on the design and conditions of use.

Which one to choose?

The right choice depends on the type of sample, how often it is accessed and the laboratory’s quality requirements:

  • Where cross-contamination is unacceptable, such as clinical cell and tissue banking, cord blood and patient samples, vapour phase, and jacketed designs in particular, stands out.
  • Where temperature stability is most critical, access is infrequent and samples are properly packaged, liquid phase remains an option in use.
  • For frequently accessed samples, the temperature distribution in the chamber, the inventory layout and the time the lid stays open should be assessed together.
  • Whichever phase is chosen, temperature and level monitoring, an alarm system and regular LN2 refilling are essential.

Inventory and packaging matter too

The rack, box and canister layout matters as much as the storage phase. Stainless steel and aluminium inventory components are durable at low temperatures and allow vapour to circulate evenly in the chamber. Freezing bags and overwrap protect the sample from both liquid nitrogen and mechanical damage.

How Cryonos Bioscience can help

For LN2 freezers operating in vapour and liquid phase, supply tanks and accessories, see our Cryogenic Products and Services page; for racks, canisters and boxes, see Cryogenic Inventory Systems. Our article on jacketed freezers and overwrap for hematopoietic stem cells complements this topic.

Let us help you choose the freezer that fits your samples and workflow. Contact us.

Sıvı azot tankı ve laboratuvar tesisatı

Cryonos: Cryogenics and Biotechnology

CONTACT

Konutkent Mah. 3035. Cadde No:74 SUİT TOWER A Blok No:12 Çankaya/ANKARA/TURKEY

+90 (555) 590 68 91

info@cryonos.com